UJI KESESUAIAN CT NUMBER IN WATER PADA PESAWAT CT SCAN MERK PHILIP DI INSTALASI RADIOLOGI RSU BMC PADANG

THE COMPATIBILITY TEST OF CT NUMBER IN WATER ON CT SCAN AIRCRAFT BRAND PHILIP IN RADIOLOGY INSTALLATION OF BMC PADANG GENERAL HOSPITAL

Authors

  • Livia Ade Nansih Universitas Baiturrahmah Padang
  • Cicilia Artitin Universitas Baiturrahmah Padang

DOI:

https://doi.org/10.38215/jtkes.v6i1.105

Keywords:

CT scan, compliance test, phantom water, tolerance value

Abstract

CT number is the value of the X-ray attenuation coefficient determined by the average energy of X-rays and the absorber atomic number is expressed by the attenuation coefficient. The role of the CT number on a CT scan machine to assess and differentiate abnormalities in human organs. Compliance tests are carried out to ensure that the CT-Scan aircraft meets radiation safety requirements and provides precise and accurate diagnostic information. This research is a quantitative research with an experimental approach, carried out at the Radiology Installation of RSU Bunda Padang in June 2022 for 5 days. The population of all CT number tests, samples at five measurement points with ROI in the center of the image, at 12 o'clock, 3 o'clock, 6 o'clock and 9 o'clock directions. Research data was taken by observing, experimenting and documenting. Data analysis was compared with PERKA BAPETEN Number 9 of 2011 concerning quality assurance of QA and QC of CT Scan aircraft. The results showed that the CT number value in the middle of the image on the first day was 4.83 HU, the second was 4.83 HU, the third was 4.14 HU, the fourth was 5.15 HU and the fifth was 5.16 HU while the average standard deviation on the first day was 3, 48 HU, the second day 3.42 HU, the third day 3.41 HU, the fourth day 3.43 HU and the fifth day 3.50 HU, the ROI uniformity results at the image and noise center show that it has exceeded the tolerance limits set by PERKA BAPETEN No. 9 of 2011.

Downloads

Download data is not yet available.

References

BAPETEN, (2011). Peraturan Kepala Badan Pengawas Tenaga Nuklir Nomor 9. Tentang Uji Kesesuaian Pesawat Sinar-X Radiologi Diagnostik dan Intervensional.

Bontrager, K.L. (2001). Radiographic Positioning and Related Anatomy textbook, Mosby A Harcourt Science Company, St. Louis London Philadelphia Sydney Toronto.

Bushberg, J.T., Seibert, J.A ., Leidholdt, E.M., Boone, J. M., (2012). The Essential Physics of Medical Imaging, third edition, Lippincott William & Wilkins, Philadelphia.

Bushong, C. S. (2000). Computed Tomography. New York: McGraw Hill Company.

Bushong, S.C., (2001). Radiologic Science for Technologists Physics, Biologic, and Protection, seventh edition, Mosby Company, USA.

Chrismawan, H. (2001). Pengaruh Tegangan Tabung Sinar-X dan Ketebalan Objek terhadap Paparan Radiasi Hambur dengan Menggunakan Dosimeter Film. Skripsi. Semarang: Universitas Diponogoro

Donis, G.I.R., Guyer, D.E., Pease, A., Fulbright, D.W. (2012). Relation of Computerized Tomography Hounsfield Unit measurement and an internal component of fresh chestnuts. Postharvest Biology and Technology.

Goldman, L.W (2007) Principle of CT Radiation Dose and Image Quality. Journal of nuclear medicine technology.

Herman, T (2009) Fundamental of Computerized Tomography, USA: Springer.

Human Health, (2012). Quality Assurance Programme For Computed Tomography, Diagnostic and Therapy Application, nineteen editions, IAEA.

Kartawiguna, D dan Rusmini. B (2017). Instrumentasi Pemindai Tomografi Komputer. Yogyakarta: Pustaka Panasena.

Manik, J., Hariyanto, B., & Abdullah, E. (2021). Analisis Kesesuaian Lapangan Kolimasi Pada Alat Digital Radiografi. Cerdika: Jurnal Ilmiah Indonesia, 1(6), 669-674.

Martono, N (2012). Metode Penelitian Kuantitatif. Jakarta: Rajawali Pers. Notoadmodjo

Mas’ull, A.R., Sutanto, H., (2014) Uji Kesesuaian CT Number pada Pesawat CT Scan Multi Slice

di Unit Radiologi Rumah Sakit Islam Yogyakarta, Youngster Physics Journal, Jurusan Fisika, Fakultas Sains dan Matematika, Universitas Diponegoro, Semarang.

Papp, J (2019) Quality Management in the Imaging Science. Sixth edition. Saint Louis: Mosby

Seeram, E. (2001). Computed Tomography Physical Principles, Clinical Applications, and Quality Control, Second Edition, W.B Saunder Company, Philadelphia.

Stern, S., H. Freier, K. Farris, A. Gantt, B. Matkovich, J. Nakasone, J. Neal, R. Scott, and M.A. Spohrer (2007) Tabulation and Graphical Summary of 2000 survey of Computed Tomography. Nationwide Evaluation of X-Ray Trends, Kentucky: CRCPD Inc.

Published

2023-01-31

How to Cite

Nansih, L. A., & Artitin, C. . (2023). UJI KESESUAIAN CT NUMBER IN WATER PADA PESAWAT CT SCAN MERK PHILIP DI INSTALASI RADIOLOGI RSU BMC PADANG: THE COMPATIBILITY TEST OF CT NUMBER IN WATER ON CT SCAN AIRCRAFT BRAND PHILIP IN RADIOLOGY INSTALLATION OF BMC PADANG GENERAL HOSPITAL. Jurnal Teras Kesehatan, 6(1), 16–21. https://doi.org/10.38215/jtkes.v6i1.105

Similar Articles

1 2 3 > >> 

You may also start an advanced similarity search for this article.